HOYA Vivinex™ Toric IOL

The glistening-free hydrophobic Vivinex™ Toric IOL offers unprecedented clarity of vision and outstanding rotational stability.

Active oxygen processing treatment

Vivinex™ is made from a novel hydrophobic acrylic, using a proprietary manufacturing process that includes a unique, active oxygen posterior surface treatment

Combined with its square edge design and smooth, regular IOL surface, it has been demonstrated to provide a low incidence of PCO in several studies.

Glistening-free hydrophobic

Vivinex™ is rated glistening-free with 97.0% of lenses demonstrating <10 glistenings per 10mm x 2mm field and showing significantly less glistenings than AcrySof® IQ SN60WF* (p<0.0001).

Outstanding rotational stability
  • Median rotation 1.1° [ range: 0.0° – 5.0° ]
  • 100% of lenses (n=103) had ≤5° of rotation from their initial axis at end of surgery through all follow up visits at 1 hour, 1 week and 6 months
HOYA Vivinex™ Toric

Download
Technical specifications
Brochures & Manuals
Description & Technical Specs

Preloaded Toric, Aspheric, Blue Light Absorbing IOL (XY1A)

Optic material Hydrophobic acrylic Vivinex™ with UV- and blue light filter
Optic design Biconvex with sharp textured optic edge Anterior: Aspheric design Posterior: Toric design
Haptic design Textured-rough haptic surface
Haptic material Hydrophobic acrylic Vivinex™ with UV- and blue light filter
Diameter (optic/OAL) 6.00 mm / 13.00 mm
Power +10.00 to +30.00 D (in 0.50 D increments)
Cylindrical Power 1.00 to 6.00 D (T2 to T9)
T2 to T3 in 0.50 D increments
T3 to T9 in 0.75 D increments
Nominal A-constant 118.9
Injector Vivinex™ iSert® preloaded
Recommended incision size 2.20mm
Front injector tip outer diameter 1.70 mm
Optimized constants Haigis a0 = -0.8394 a1 = 0.2023 a2 = 0.2272
Hoffer Q pACD = 5.7058
Holladay 1 sf = 1.9432
SRK/T A = 119.219 (VivinexTM iSert® preloaded)

The A-constant is presented as a starting point for the lens power calculation. When calculating the exact lens power, it is recommended that calculations be performed individually, based on the equipment used and operating surgeon’s own experience.

Optimized Optical constants Source: https://iolcon.org Calculated from 910 clinical results based on caucasian patients for Vivinex™ Model XY1/XC1 as of August 22, 2019.

Model XY1A Cylinder Power
at IOL Plane
Cylinder Power
at Corneal Plane5
T2 1.00 D 0.69 D
T3 1.50 D 1.04 D
T4 2.25 D 1.56 D
T5 3.00 D 2.08 D
T6 3.75 D 2.60 D
T7 4.50 D 3.12 D
T8 5.25 D 3.64 D
T9 6.00 D 4.17 D

5Based on an average pseudophakic human eye.